User`s guide
Acquiring PSD Data with Standard Instrument Settings (.BIC) Files
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 7-25
7
2. If the PSD Acquisition Settings control page (Figure 7-8) 
is not displayed, select PSD Acquisition from the View 
menu.
Figure 7-8 PSD Acquisition Settings Control Page
NOTE: The Precursor Ion Selector in PSD mode is the 
same parameter as the Timed Ion Selector in Reflector 
mode. If you change the Precursor mass used by the 
Precursor Ion Selector in PSD mode, the mass 
specified for the Timed Ion Selector in Reflector mode 
also changes.
Setting voltages
and external
calibration for the
precursor
spectrum
3. For optimum mass accuracy, change the Accelerating 
Voltage and Grid Voltage% (on the Instrument Settings 
control page) to the values you used to acquire the 
precursor spectrum and to generate the calibration.
NOTE: If you use different Accelerating Voltage, Grid 
Voltage, or Delay Time to acquire the precursor 
spectrum and generate the calibration than you use to 
acquire the fragment spectra, the software 
compensates for the different values. However, using 
different voltages may not yield optimum mass 
accuracy.