Chapter 6 Acquiring Mass Spectra
6-72 PerSeptive Biosystems
6
Internal/External 
Calibration
Calibration type to use for this row. Click the down arrow and 
select one of the following:
• Blank—No calibration applied during processing. 
Acquisition calibration (calibration specified in .BIC) is 
persisted.
• External—Applies the constants in the .CAL file specified to 
the data file acquired in this row.
• Internal—Applies the constants in the .CAL file if specified, 
generates new calibration constants using the reference 
masses specified in the .SET file, and applies the new 
constants to the data file acquired in this row. For more 
information, see Section 6.7.7.4, Internal Standard 
Calibration Considerations. 
• Internal-Update—Applies the constants in the .CAL file, if 
the specified file exists, generates new calibration constants 
using the reference masses specified in the .SET file, 
applies the new constants to the data file acquired in this 
row, and saves the new constants in the .CAL specified. For 
more information, see Section 6.7.7.3, Performing Close 
External Calibration.
For information, see:
• Section 6.7.1, Understanding Settings, 
Macros, and Calibration
• “Creating calibration (.CAL) files” on page 6-66
NOTE: Calibration specified in the Sequence Control Panel is 
performed after calibration performed by the .BIC file used to 
acquire the data file. Calibration performed by the Sequence 
Control Panel overrides the calibration performed by the .BIC 
file.
NOTE: If the .BIC file for the row specifies a save/accumulation 
mode that creates more than one data file, calibration is 
performed on all data files created.
Table 6-4 Run List Parameters (Continued)
Field Description