User`s guide
Acquiring in Automatic Mode from the Instrument Control Panel
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 6-53
6
NOTE: If no minimum or maximum signal intensity 
criteria is specified, the laser is not adjusted. The 
mid-range laser setting specified in the Automatic 
Control dialog box is used.
• Within range—The system saves the data file if 
Signal-to-Noise is not enabled, or continues with 
step 2 if Signal-to-Noise is enabled.
2. If Signal-to-Noise is enabled in the Spectrum Acceptance 
Criteria dialog box, the system evaluates the 
signal-to-noise ratio for each spectrum.
If the signal-to-noise ratio is:
• Equal to or greater than the ratio set in the 
Spectrum Acceptance Criteria dialog box—The 
system checks resolution as described in step 3, if 
it is enabled. If resolution is not enabled, the 
system saves the data file.
• Less than the ratio set in the Spectrum 
Acceptance Criteria dialog box—The system 
increases the laser setting one laser step as 
specified in instrument settings, and acquires a 
new spectrum.
3. If Resolution is enabled in the Spectrum Acceptance 
Criteria dialog box, the system evaluates the resolution for 
each spectrum.
If the resolution is:
• Equal to or greater than the resolution set in 
the Spectrum Acceptance Criteria dialog box—
The system saves the data file.
• Less than the resolution set in the Spectrum 
Acceptance Criteria dialog box—The system 
acquires a new spectrum.