Acquiring in Automatic Mode from the Instrument Control Panel
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 6-51
6
2. The system sets the laser to the minimum setting 
specified in the Automatic Control dialog box and 
acquires a spectrum. 
If the signal intensity is:
• Too low with the laser at minimum—The system 
continues with step 3.
• Too high with the laser at minimum—The 
system begins acquiring in Acquisition mode.
• Within range—The system begins acquiring in 
Acquisition mode.
3. The system sets the laser halfway between the minimum 
and maximum and acquires spectra.
For the first set of Shots/Spectrum, minimum and 
maximum laser settings are taken from the Automatic 
Control dialog box. For subsequent Shots/Spectrum, 
minimum and maximum laser settings are determined 
as described in step 4.
4. If the signal is too low, the current setting is used as the 
new minimum setting. If the signal is too high, the current 
setting is used as the new maximum setting.
5. The system repeats step 3 and step 4 until either the 
signal falls between the specified maximum and minimum 
values, or the difference between the maximum and 
minimum laser intensity in step 3 is smaller than the laser 
step size specified in the Automatic Control dialog box.