User`s guide
Acquiring in Automatic Mode from the Instrument Control Panel
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 6-43
6
6.6.3 Automatically Acquiring, 
Evaluating, and Saving Spectra
To automatically acquire a spectrum:
1. In the Instrument Control Panel, open or create an 
instrument settings file with the appropriate parameters. 
See Section 6.6.2, Setting Instrument Settings for 
Automatic Control Mode, for information.
2. Set Data Storage parameters as described in “Setting 
Data Storage parameters” on page 6-14.
NOTE: You must specify a file name in the Data 
Storage control page before acquiring in Automatic 
Control mode.
3. In the Manual Laser/Sample Positioning control page, 
select the Active Position to analyze. 
4. To start acquiring, select Start Acquisition from the 
Acquisition menu, or click  .