User`s guide
Acquiring in Automatic Mode from the Instrument Control Panel
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 6-41
6
Table 6-3 Spectrum Acceptance Criteria Parameters
Parameter Description
Acceptance Criteria
Minimum Signal 
Intensity
(spectrum acceptance 
and laser adjustment)
Sets the minimum signal intensity accepted for the most 
abundant peak (local base peak) within the mass range of 
interest.
Also used to determine laser adjustment. When the 
system adjusts the laser, it checks that the signal intensity 
is above the minimum specified. If signal intensity is not 
above the minimum, the system increases the laser 
intensity. If signal intensity is not above the minimum 
when the laser is at the highest setting specified in 
Automatic Control, the system moves to the next search 
pattern position.
Maximum Signal 
Intensity
(spectrum acceptance 
and laser adjustment)
Sets the maximum signal intensity for the most abundant 
peak (local base peak) within the mass range of interest.
Also used to determine laser adjustment. When the 
system adjusts the laser, it checks that the signal intensity 
is below the maximum specified. If signal intensity is not 
below the maximum, the system decreases the laser 
intensity. If signal intensity is not below the maximum 
when the laser is at the lowest setting specified in 
Automatic Control, the system moves to the next search 
pattern position.
Signal-to-Noise
(spectrum acceptance 
and laser adjustment)
Enables or disables the signal-to-noise filtering. Type the 
signal-to-noise ratio threshold. Spectrum is rejected if the 
base peak has a ratio below this value.
The base peak in the specified Mass Range is used to 
calculate signal. The noise is calculated over the Baseline 
Range specified (see below).
Continued