User`s guide
Chapter 6 Acquiring Mass Spectra
6-30 PerSeptive Biosystems
6
5. Type in up to four Mass/Charge values for which to 
calculate resolution.
6. For each Mass/Charge, enter the window for calculation 
(± AMU).
NOTE: To label isotopes, set the ± AMU value low 
enough to prevent the calculation windows for each 
isotope peak from overlapping. If the calculation 
windows overlap, only the highest peak is labeled. If 
you set too low, the peak of interest may not be labeled. 
However, if you set ± AMU too low, the peak of interest 
may not appear in the window, and resolution will not 
be calculated.
7. Click OK.
The peaks are labeled with (RXXXX) next to the peak 
mass, where XXXX is the resolution (Figure 6-13). To 
turn off resolution labels, see “Labeling spectrum 
peaks” on page 6-28.
NOTE: If you do not apply peak detection parameters 
after acquisition is complete, resolution labels are not 
displayed.