User`s guide
Chapter 6 Acquiring Mass Spectra
6-26 PerSeptive Biosystems
6
• Signal intensity of reference and analyte peaks is in 
the same range—Signal intensity of reference and 
analyte peaks should be of similar magnitude. If they 
are not in the same range, you may see weak analyte 
peaks with poor signal-to-noise ratio while the internal 
standard is adequate, or saturated internal standard 
peaks with adequate analyte peaks.
• No contaminants present—Unresolved contaminants 
can affect peak shape.
• Two internal reference masses that bracket the 
mass range of interest—Best results are obtained 
when masses above and below the mass of interest are 
correctly identified.
NOTE: Although not required, using more than two 
reference masses can help to minimize the effects of 
an asymmetrical reference peak. 
• Internal reference masses are within the same 
range as analyte masses—Best results are obtained 
when masses are in the same range. For example, 
reference masses of 904 and 1,570 are adequate when 
evaluating an analyte mass of 1,296 Da.
During mass
calibration and
peak detection
When calibrating mass (described in the
 Data Explorer 
Software User’s Guide, 
Section 5.3, Manual Calibration) and 
detecting peaks (described in the
 Data Explorer Software 
User’s Guide, 
Section 3.2, Peak Detection), the following 
factors contribute to the accuracy of mass measurements:
• Use Centroid mass instead of Apex mass—Before 
calibrating, set peak labels to Centroid. Centroid mass 
is the best representation of peak position. Apex mass 
represents the highest data point.
• Set Centroid Percent to 50%—Lower settings can 
include noise or contaminant peaks. Higher settings 
may not include adequate data to ensure accurate 
mass assignment.










