User`s guide
Chapter 6 Acquiring Mass Spectra
6-16 PerSeptive Biosystems
6
During
acquisition
During acquisition:
• The Live/Current Spectrum trace in the Spectrum window 
updates to display the spectrum that results from each 
laser shot.
NOTE: If your system includes an oscilloscope, the 
Current Spectrum does not display a trace until 
acquisition is complete. The spectra that result from 
each laser shot are displayed on the oscilloscope 
screen.
• The system averages all spectra acquired since you 
started acquisition. 
When acquisition is complete, the software displays the 
averaged spectrum in a Current trace in the Spectrum window. 
Evaluating data
As data is acquired, observe the trace in the Spectrum 
window:
1. Examine the number at the top of the right axis in the 
Current Spectrum trace to determine if signal is 
saturated (Figure 6-4). Signal with counts greater than 
64,000 may be saturated. Adjust the laser intensity as 
needed.
For information, see “Signal saturation” on page 5-59.
2. Use the Resolution calculator or the Signal-To-Noise 
calculator on the Current trace to determine if data is 
acceptable. For more information see Section 6.5, 
Evaluating Data in the Instrument Control Panel.
Hint: You can accumulate traces to improve resolution 
and signal-to-noise ratio. For more information, see 
Section 6.2.2, Manually Accumulating Spectra 
from Multiple Acquisitions.