User`s guide
Optimizing Instrument Settings Parameters
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 5-81
5
Figure 5-20 Low Mass Gate Spike 
The spike occurs at a time that corresponds to just below the 
Mass for the Low Mass Gate entered in the .BIC file. For 
example, if the Mass is set to 400 Da, you would see the spike 
at approximately the time that corresponds to 370 Da.
You can identify the Low Mass Gate spike by its sharp lift-off, 
its narrow width, and the noise as it returns to baseline. 
Eliminating Low
Mass Gate spike
when acquiring
You can eliminate the Low Mass Gate spike from the data by 
setting the Mass for acquisition slightly higher than Low Mass 
Gate Mass.
Eliminating Low
Mass Gate spike
after acquiring
You can eliminate the Low Mass Gate spike after you acquire 
by using the Truncate Spectrum function in the Data Explorer 
software to eliminate the Low Mass Gate spike. For 
information, see the
 Data Explorer Software User’s Guide
,
Section 5.9, Truncating a Spectrum.
Low Mass Gate spike