User`s guide
Chapter 5 Optimizing Instrument Settings
5-80 PerSeptive Biosystems
5
When to use
Turn on Low Mass Gate when:
• Analyzing masses greater than 2,000–3,000 Da
• Matrix signal is more intense than the sample signal
Optimum setting
for starting mass
You will need to experiment with the optimum setting for the 
starting mass (threshold) for Low Mass Gate. As a general 
guideline:
• If you are analyzing peptides in the 400 to 2,000 Da 
range, set the Low Mass Gate ~350 Da
• If you are analyzing peptides in the 2,000 to 8,000 Da 
range, set the Low Mass Gate ~500 Da
• If you are analyzing samples in the 10,000 to 30,000 Da 
range, set the Low Mass Gate ~3,000 Da
When not to use
Do not use Low Mass Gate if:
• You are analyzing peaks in the same mass range as 
matrix peaks
• You are using a matrix peak to calibrate
NOTE: Low mass gate is automatically disabled when you 
change the instrument mode to PSD.
Low Mass Gate
spike
When Low Mass Gate is turned on, a characteristic spike may 
occur in the spectrum (Figure 5-20). This spike is typically only 
seen in Reflector mode.
NOTE: Systems with high current detectors (Voyager-DE 
systems with serial number 1128 and later; Voyager-DE 
PRO systems with serial number 6007 and later; 
Voyager-DE STR systems with serial numbers 4112, 4113, 
and 4116 and later) do not typically exhibit this spike. Older 
systems (with a dual channel plate detector in Linear mode) 
may exhibit a 50 to 1,000 mV spike.










