User`s guide
Optimizing Instrument Settings Parameters
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 5-79
5
5.4.4.4  Setting Shots/Spectrum
Overview
A higher number of Shots/Spectrum can improve 
signal-to-noise ratio, and also improve the dynamic range of 
the acquisition. However, acquiring a higher number of 
Shots/Spectrum increases acquisition time and file size.
When random noise is present in a spectrum, the 
improvement in signal-to-noise ratio is approximately 
proportional to the square root of the number of 
Shots/Spectrum taken.
Manually
accumulating to
override limit
The maximum number of Shots/Spectrum is determined by 
the digitizer on your system (see “Shots/Spectrum” on 
page 5-19). You can acquire a greater number of 
Shots/Spectrum than the maximum number of 
Shots/Spectrum noted above by manually accumulating 
spectra. For more information, see Section 6.2.2, Manually 
Accumulating Spectra from Multiple Acquisitions.
5.4.4.5  Setting Low Mass Gate
Description
Low Mass Gate suppresses matrix peaks that can interfere 
with ion detection which can saturate the detector as laser 
intensity increases. This saturation effect is less pronounced 
on systems with Linear High Current Detectors. For serial 
numbers of systems that include High Current Detectors, see 
page 1-23 and page 1-36.
The detector is not turned on until ions greater in mass than 
the mass you enter pass the detector. 
NOTE: When Low Mass Gate is enabled, the mass range 
below the Low Mass Gate starting mass is still detected, but 
detector sensitivity below this mass is decreased by 100 to 
1,000 fold. If strong matrix signal is present below the Low 
Mass Gate, you can use it for calibration if desired.










