User`s guide
Chapter 5 Optimizing Instrument Settings
5-68 PerSeptive Biosystems
5
Optimizing Delay
Time
NOTE: If you are analyzing a broad mass range and need 
optimum resolution in all mass regions, you may need to 
acquire smaller portions of the mass range and set the 
Delay Time to optimize resolution for each mass range.
To optimize Delay Time:
1. Open a standard .BIC file for the mass range you are 
acquiring. For more information, see “Selecting a .BIC 
file” on page 5-8.
2. Acquire a spectrum and observe the resolution. For 
information, see Section 6.5.2, Calculating Mass 
Resolution.
If the resolution is not acceptable, increase the Delay 
Time by 100 nsec (Linear mode) or 50 nsec (Reflector 
mode), acquire a spectrum and observe. Leave the Grid 
Voltage% setting unchanged.
If the signal is saturated at the higher Delay Time 
setting, decrease the laser intensity and acquire a new 
spectrum. If the signal decreases at the higher Delay 
Time setting, increase the laser and acquire a new 
spectrum.
Hint: If you are analyzing a mixture and the resolution 
of the peak of interest is not acceptable, observe the 
resolution of a higher mass peak and a lower mass 
peak. If resolution is better on the higher mass peak, 
decrease the Delay Time (by 100 nsec for Linear mode or 
50 nsec for Reflector mode). If resolution is better on the 
lower mass peak, increase the Delay Time (by 100 nsec 
for Linear mode or 50 nsec for Reflector mode). Collect a 
spectrum and observe. Leave the Grid Voltage% setting 
unchanged.