User`s guide
Chapter 5 Optimizing Instrument Settings
5-48 PerSeptive Biosystems
5
Default Number of
Data Points/Bin
Size settings
The software uses the following default digitizer settings:
• Number of Data Points—100,000
• Bin size—Smallest possible Bin size to accommodate 
the mass range selected.
Number of Data Points and Bin size are dependent 
values, if you change one value the other changes. For 
more information, see “Mass Range (Da)” on page 5-19.
Overriding
horizontal
digitizer settings
You can override the default digitizer settings:
1. In the Instrument Settings control page, click 
Mode/Digitizer.
2. Click Linear or Reflector Digitizer, and select the 
digitizer parameter to optimize:
• Number of Data Points—Allows you to select the 
memory size available for sampling. Increasing 
this value decreases the Bin size. Decreasing this 
value increases the Bin size.
• Bin Size—Allows you to specify the Bin size 
(nanoseconds). For information, see “Effects of 
adjusting Bin size” on page 5-49.
3. Change the Bin size to improve resolution as needed:
NOTE: Setting the Bin size lower than the default calculated 
by the system may result in a larger data file size.
Mode Digitizer Bin Size
Linear 1 nsec
NOTE: A Bin size setting below 1 nsec does 
not typically improve observed mass 
resolution in Linear mode
Reflector 0.5 or 1 nsec