User`s guide
Chapter 5 Optimizing Instrument Settings
5-40 PerSeptive Biosystems
5
Grid Voltage% Linear/ 
Reflector
Critical parameter with optimum value for 
maximum resolution.
Digitizer Bin size 
(nanoseconds)
Linear/ 
Reflector
Decreasing improves resolution.
NOTE: Decreasing Bin size increases file 
size.
Digitizer Input 
Bandwidth (not 
available with Signatec 
digitizers)
Linear/ 
Reflector
Decreasing may improve signal-to-noise 
ratio.
Increasing may improve the resolution.
Accelerating Voltage Linear Increasing improves sensitivity and 
resolution, but is limited by other factors 
such as the digitization rate.
Reflector Decreasing to 10,000 to 15,000 V can 
improve resolution for compounds 
<2,000 Da.
Laser intensity Linear/ 
Reflector
Increasing improves signal-to-noise ratio.
Increasing too high increases 
fragmentation, and resolution is sacrificed 
due to saturation.
Decreasing too low decreases sensitivity.
Shots/
Spectrum
Linear/ 
Reflector
Increasing improves signal-to-noise ratio.
Parameter Mode Impact