Instrument Settings Parameter Descriptions
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 5-35
5
Save Conditions
(continued)
• Save first spectrum to pass acceptance criteria—
The first spectrum that meets the selected 
acceptance criteria is saved. Acquisition is performed 
on each search pattern position* until a spectrum 
passes or until the end of the search pattern is 
reached. If no spectra pass acceptance criteria, no 
data file is saved. All search pattern positions may 
not be analyzed. 
Example: Number to Acquire is automatically set to 
1, Number of positions in .SP=20, the number of 
positions analyzed is equal to the position number in 
which the first passing spectrum is found. One .DAT 
created. If no spectra meet criteria, no .DAT created.
• Save best—Automatic Sample Positioning is 
required with this mode. Acquisition is performed on 
all search pattern positions. All Acceptance Criteria 
are evaluated, and the passing spectrum with the 
best Signal-to-Noise result is saved. If no spectra 
pass acceptance criteria, no .DAT created.
NOTE: In this mode, acquisition moves to the next search 
pattern position for every acquisition, even if acceptance 
criteria pass.
Example: Number to Acquire is automatically set to 
1, Number of positions in .SP=20, the number of 
positions analyzed=20. One .DAT created containing 
the best spectrum acquired. If no spectra meet 
criteria, no .DAT created.
* If Use Automated Sample Positioning is disabled, acquisition is performed on the positions you 
manually select.
Table 5-10 Automatic Control Parameters—Spectrum Accumulation (Continued)