User`s guide
Chapter 5 Optimizing Instrument Settings
5-34 PerSeptive Biosystems
5
Save Conditions
(continued)
• Save all spectra that pass acceptance criteria— 
Each spectrum that meets the specified acceptance 
criteria is saved in a separate data file. Acquisition is 
performed on the same search pattern position* until 
Acceptance Criteria fail. The number of .DAT files 
created is equal to the number of spectra that pass 
acceptance criteria. Acquisition continues until the 
Number of Spectra to Acquire have been acquired or 
the end of the search pattern is reached. All search 
pattern positions may not be analyzed. 
NOTE: In this mode, acquisition moves to the next search 
pattern position only if acceptance criteria fail.
Example: If Number to Acquire=5, Number of 
positions in .SP=7, the number of positions analyzed 
and the number of .DAT created is determined by 
whether acceptance criteria fail:
• If all fail, 7 positions analyzed (total number of positions 
in .SP), no .DAT created
• If the first position fails, but all remaining pass, 
2 positions analyzed, 5 .DAT created
• If the first and second positions fail, but all remaining 
pass, 3 positions analyzed, 5 .DAT created, and so on
• If acceptance criteria never fail, 1 search pattern 
position analyzed
* If Use Automated Sample Positioning is disabled, acquisition is performed on the positions you 
manually select.
Table 5-10 Automatic Control Parameters—Spectrum Accumulation (Continued)
Parameter Description