User`s guide
Instrument Settings Parameter Descriptions
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 5-33
5
Table 5-10 Automatic Control Parameters—Spectrum Accumulation
Parameter Description
Spectrum Accumulation
Number to Acquire Number of spectra to save or accumulate. This field is 
dependent on the condition selected in the conditions 
field. For example, if you select Save First or Save Best, 
the number to acquire is restricted to one.
Save Conditions The following Save conditions create one or more data 
files that contain one spectrum each:
• Save all spectra—All spectra are saved. One 
spectrum per search pattern position* is acquired. 
The number of .DAT files created is equal to the 
Number of Spectra to Acquire, even if the number of 
positions in the search pattern is larger. All search 
pattern positions may not be analyzed. Acceptance 
criteria are not used.
Example: If Number to Acquire=5, Number of 
positions in .SP=20, 5 positions analyzed, 5 .DAT 
created. If Number to Acquire=25, Number of 
positions in .SP=20, 20 positions analyzed, 20 .DAT 
created. 
* If Use Automated Sample Positioning is disabled, acquisition is performed on the positions you 
manually select.