Chapter 5 Optimizing Instrument Settings
5-30 PerSeptive Biosystems
5
Table 5-8 Advanced Parameters
Parameter Description
Mirror to Accelerating 
Voltage Ratio
(not available in Linear 
mode)
Specifies the ratio between the Mirror Voltage and the 
Accelerating Voltage in Reflector mode, to adjust the 
voltage of the mirror so that it is slightly higher than the 
Accelerating Voltage. A higher voltage is needed at the 
mirror to reflect ions. If the voltage at the mirror is equal to 
the Accelerating Voltage, ions will pass the mirror.
To reflect intact ions, the Mirror to Accelerating Voltage 
Ratio must be close to or greater than 1. Maximum value 
is:
30,000V / Accelerating Voltage
NOTE: This parameter influences the tuning 
characteristics and default calibration of the instrument. 
Do not change this setting unless instructed to do so by a 
PerSeptive Biosystems Technical Representative. 
Changing this value significantly alters the optimum 
settings of Grid Voltage% and Delay Time.
For information on the voltage applied to the mirror in PSD 
mode, See Section 7.1, Overview of PSD Analysis. 
Timed Ion Selector 
(with mass)
Allows analysis of the mass of interest by deflecting other 
ions. This parameter is typically only used in PSD mode, 
but is available in Reflector mode for advanced 
applications. See Section 7.1, Overview of PSD Analysis, 
and Section 7.4.2, Observing the Effects 
of Precursor Ion Selector, for more information.
NOTE: The Timed Ion Selector in Reflector mode is the 
same parameter as the Precursor Ion Selector in PSD 
mode. If you change the mass for Timed Ion Selector in 
Reflector mode, the Precursor mass used by the 
Precursor Ion Selector in PSD mode also changes.