User`s guide
Instrument Settings Parameter Descriptions
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 5-19
5
Shots/Spectrum Determines the number of laser shots that each spectrum 
will contain. For more information, see Section 5.4.4.4, 
Setting Shots/Spectrum.
The maximum number of Shots per Spectrum for each 
digitizer are as follows:
• Signatec—256
• LeCroy—1,000
• Tektronix—10,000
NOTE: The actual number of times the laser fires may be 
greater than the number of Shots/Spectrum specified, due 
to the limited averaging speed of the digitizer. For 
example, if 250,000 data points are recorded, every other 
laser shot is skipped by the digitizer. In any case, the 
specified number of Shots/Spectrum will be stored in the 
data file.
Spectrum Acquisition
Mass Range (Da) Determines the mass range collected.
Valid ranges are:
• Start Mass—0 to End Mass.
• End Mass—Start Mass to End Mass.
For information on the default digitizer settings used with 
the mass range specified, see “Default Number of Data 
Points/Bin Size settings” on page 5-48.
Table 5-4 Instrument Settings Parameters (Continued)
Parameter Description