User`s guide
Instrument Settings Parameter Descriptions
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 5-17
5
Voltages
Accelerating Voltage Voltage applied to the first stage ion source. Valid range is 
0 to 25,000 V. 
For information on settings for different mass ranges, see 
Section 5.4.4.2, Setting Accelerating Voltage.
NOTE: The calibration of the mass scale changes 
significantly when you change the Accelerating Voltage. 
Default calibration adjusts for these changes. However, 
you will observe more accurate calibration if you use an 
external calibration (.CAL) file generated with the same 
Accelerating Voltage. See the Data Explorer Software 
User’s Guide, Section 5.3.2, Manually Calibrating, for 
more information.
Grid Voltage% Voltage applied to the grid above the sample plate. Valid 
range is determined by the Accelerating Voltage. Optimize 
in conjunction with Delay Time (described on page 5-18).
See Section 5.3.2, Understanding Grid Voltage%, and 
Section 5.4.3.5, Optimizing Grid Voltage%, for more 
information.
Table 5-4 Instrument Settings Parameters (Continued)
Parameter Description