User`s guide
Chapter 3 Preparing Samples
3-58 PerSeptive Biosystems
3
NOTE: You can correct for systematic errors introduced by 
a sample preparation device by aligning the sample plate in 
the Voyager Workstation. See Section 2.7, Aligning the 
Sample Plate.
If you are using different sample preparation devices, you 
can compensate for different systematic errors by creating 
different .PLT files and sample plate alignments for each 
system.
Effect of plate type
on area available
for analysis
When creating search pattern files, consider the plate types 
used:
• If you are using laser-etched position plates, the area 
available for analysis is limited to the well or laser-etched 
position size (2,540 µm on a 100-well plate).
• If you are using plates with polished blank surfaces, the 
area available for analysis is determined by the proximity 
of other positions specified. Approximately half the 
distance between the X,Y coordinates of two adjacent 
positions is available for analysis of each position. For 
example, if the center X,Y coordinates of the two 
positions are 5,080 µm apart (plus positional tolerance, 
see below), approximately 2,540 µm around the 
X,Y coordinates is available for analysis of each position 
(Figure 3-12 on page 3-60).