User`s guide
Sample Plate Types Supported
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 3-57
3
Guidelines for
creating .PLT files
Note the following:
• The diameter of a position on a plate is determined by the 
plate used:
• If you are using welled or laser-etched plates, the 
position diameter is determined by the well or 
laser-etched position size (2,540 µm on a 100-well 
plate).
• If you are using plates with polished blank 
surfaces, the position diameter is determined by 
the size of the physical sample spot delivered by 
manual application or automated sample 
preparation device.
• When creating a .PLT file, specify absolute X,Y 
coordinates that accommodate the sample diameter, but 
do not allow samples to overlap.
• After you create a .PLT file, create an .SP file to 
accommodate the sample diameters defined in the .PLT 
file. The DEFAULT.SP file is set to analyze 2,540-µm 
diameter positions, corresponding to standard sample 
positions on a 100-well plate.
NOTE: For optimum mass accuracy, do not spot sample on 
outer rows of the plate. See Figure 3-11 on page 3-56.
Guidelines for
creating .SP files
When creating .SP files to analyze sample positions defined in 
custom .PLT files, note that the area available for analysis 
around the absolute X,Y coordinate of the position specified in 
the .PLT file is determined by three factors: 
• Plate being used
• Tolerances and non-systematic errors
• Search pattern radius
Search pattern files are described in detail in Section 6.6.4, 
Search Pattern Files.