Loading Sample Plates in the Mass Spectrometer
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 3-41
3
3.4.1 Ejecting the Sample Holder
In the Voyager Instrument Control Panel, select Eject from the 
Sample Plate menu. The Load/Eject dialog box is displayed.
. 
Figure 3-4 Load/Eject Sample Plate Dialog Box
The following also occurs:
• A Load Status dialog box is displayed during the ejection 
sequence which displays hardware status.
• High voltage is turned off. 
• The video monitor displays the sample plate moving.
• The sample holder moves out of the main source 
chamber, out of the sample loading chamber, and is 
ejected from of the instrument.
NOTE: Do not leave the sample holder in the Eject position 
for more than a few seconds. Minimize the time the 
instrument is exposed to the atmosphere to reduce the time 
needed to reach high vacuum when you insert a new plate.