Loading Samples on Sample Plates
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 3-31
3
3.2.2 Loading Samples 
(Dried Droplet Application)
NOTE: If you are loading a plate for acquisition or 
automated sample positioning in Automatic Control mode, 
use a laser-etched or welled sample plate to provide a 
reference point for sample application.
When to use
Use this technique for most applications, when sample 
concentration is >0.1 pmol/ µl.
Location of
standards
For optimum mass accuracy with welled plates or plates with 
laser-etched indentations, place standards in wells or 
indentations adjacent to the samples for which you are 
calibrating. Do not use the outer wells on the sample plate. 
when using flat or polished blank plates.
Loading samples
1. Create a log sheet listing sample position and sample 
name. Appendix D, Log Sheets, contains a master 
sheet that you can copy and use.
NOTE: For optimum mass accuracy, do not use the 
outer rows of the sample plate, unless using internal 
standards.
2. Using a clean pipette tip for each new sample, load the 
following:
• If you premixed samples—Load 0.5 to 2 µl of 
sample/matrix solution on the appropriate position. 
• If you are mixing sample and matrix on the 
sample plate—See “Mixing sample and matrix on 
the sample plate” on page 3-27.
• If you prepared samples with ZipTips—Load 0.5 
to 2 µl of sample/matrix solution on the appropriate 
position. 