User`s guide
Chapter 3 Preparing Samples
3-16 PerSeptive Biosystems
3
You can load polymer samples on sample plates in two ways:
• Thin layer polymer method (yields even response, 
but provides adequate sample response for only 
10 laser shots)—Load 0.3 µl sample/matrix solution in 
one sample position.
• Thick layer polymer method (yields less even 
response, but provides adequate sample response 
for 50 to 100 laser shots)—Load 2 to 3 µl sample/matrix 
solution in one sample position by placing the pipet tip in 
the sample position and slowly expelling while keeping 
the tip in contact with the sample plate.
NOTE: In general you should not touch the pipette tip 
to the sample plate surface. During the thick layer 
method it is necessary to touch the tip to the sample 
plate when using THF as the solvent, in order to slowly 
expel the sample onto the sample plate.