User`s guide
Aligning the Sample Plate
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 2-41
2
Overview of video
monitor display
The following examples show how the sample positions and 
laser spot may be displayed on the video monitor during 
sample plate alignment.
Example Description
Figure 2-23 Ideal Sample Position
Alignment
Ideal sample position 
alignment—The center of 
the sample position is 
aligned with respect to the 
laser spot. 
You mark the outline of 
this alignment on a 
transparency to use to 
calibrate other sample 
positions during 
alignment.
Figure 2-24 Sample Position Misalignment
Sample position 
misalignment—The 
center of the sample 
position is not aligned with 
respect to the laser spot. 
You adjust misaligned 
sample positions to ideal 
alignment.
Figure 2-25 Camera Misalignment
Camera misalignment—
The laser spot is not 
centered on the video 
monitor. The camera 
requires adjustment.
Do not perform the sample 
plate alignment procedure.
Call PerSeptive 
Biosystems Technical 
Support.
Laser spot
Perimeter of 
sample position
Outline of 
i
deal
 position alignment
Sample position
Laser spot
Laser spot