Selecting the waveform update mode
Background The display update mode is switched automatically or manually according to the
horizontal scale.
Main mode Updates the whole displayed waveform at once. The main mode is automatically
selected when the horizontal scale (timebase) is fast.
Horizontal scale ≤100ms/div
Trigger All modes available
Roll mode Updates and moves the waveform gradually from the right side of the display
to the left. The Roll mode is automatically selected when the horizontal scale
(timebase) is 50ms or slower.
When in the Roll mode, an indicator appears at the bottom of the display. When
in roll mode the record length is 2M (1 channel) or 1M (2 channel).
Main mode Roll mode
Timebase ≥50ms/div (≤1.25MS/s)
Trigger Auto mode only
Selecting the
Roll mode
manually
1. Press the Horizontal menu key.
MENU
2. Press Roll. The horizontal scale automatically
becomes 50ms/div and the waveform starts
scrolling from the right side of the display (If the
oscilloscope is already in the Roll mode, there
will be no change).
Roll