Average Multiple data is averaged to form a
waveform. This mode is useful for
drawing a noise-free waveform. To select
the number, press Average repeatedly.
Average number: 2, 4, 8, 16, 32, 64,
128, 256
Peak detect To activate the Peak detect mode, press
Peak-Detect. Only the minimum and
maximum value pairs for each acquisition
interval (bucket) are used. This mode is
useful for catching abnormal glitches in
a signal.
Peak detect
effect using the
probe comp.
waveform
1. One of the probe compensation
waveforms can demonstrate the peak
detection mode. Connect the probe to
the probe compensation output.
2. Press the Utility key.
Utility
3. Press ProbeComp.
ProbeComp
Menu
4. Press Wave Type and select the
waveform.
Wave Type
5. Press the Autoset key. The oscilloscope
positions the waveform in the center of
the display.
Autoset
6. Press the Acquire key
Acquire