User manual
69
Edit: NoGo Violation Conditions
Procedure 1. Press the Utility key.
Utility
2. Press the More key.
More
3. Press the Go-NoGo Menu key.
Go - NoGo
Menu
4. Press Violating repeatedly to select the
NoGo conditions.
Violating
Stop
Stop Stops the test when the NoGo conditions have been met.
Continue The tests continue even when the NoGo conditions have been met.
Edit: Template (boundary)
Background The NoGo template sets the upper and lower amplitude boundary. Two methods
are available: Min/Max and Auto.
Min/Max Selects the upper boundary (Max) and lower boundary
(Min) as separate waveforms, from the internal memory.
The upper boundary is saved to Ref A, the lower
boundary is saved to Ref. B.
Advantage: The template shape and distance
(allowance) between the source signal are fully
customizable.