The waveform does not update (frozen).
Press the Run/Stop key to unfreeze the waveform. See page 50 for details. For trigger setting details,
see page 96. If this does not help, press the CH key. If the signal still does not appear, press the
Autoset key.
The probe waveform is distorted.
You might need to compensate the probe. For details, see page 135. Note that the frequency accuracy
and duty factor are not specied for probe compensation waveforms and therefore it should not be
used for other reference purposes.
Autoset does not catch the signal well.
The Autoset function does not catch signals well under 30mV or 2Hz.
Please operate the oscilloscope manually. See page 48 for details.
I want to clean up the cluttered panel settings.
Recall the default settings by pressing the Save/Recall key Default Setting. For default setting
contents, see page 45.
The saved display image is too dark on the background.
Use the Inksaver function which reverses the background color. For details, see page 121.