135
7. The Channel1 calibration will
complete in less than 5 minutes.
Ch1 calibration 1/3
8. When î‚¿nished, connect the
calibration signal to the Channel
2 input and repeat the procedure.
Done!!
9. When the calibration is complete the display will go back to the previous state.
Probe Compensation
Procedure 1. Connect the probe between the Channel1 input and the probe compensation
output (2Vp-p, 1kHz square wave) on the front panel. Set the probe voltage
attenuation to x10.
2. Press the Utility key.
Utility
3. Press ProbeComp.
ProbeComp
Menu