Safety Guidelines
General
Guideline
Make sure the BNC input voltage does not exceed 300V peak.
Never connect a hazardous live voltage to the ground side of the BNC
connectors. It might lead to re and electric shock.
Do not place heavy objects on the oscilloscope.
Avoid severe impact or rough handling that may damage the oscilloscope.
Avoid discharges of static electricity on or near the oscilloscope.
Use only mating connectors, not bare wires, for the terminals.
Do not block the cooling fan vent.
Do not perform measurements at power sources and building installation sites
(Note below).
The oscilloscope should only be disassembled by a qualied technician.
(Measurement categories) EN 61010-1:2001 species the measurement
categories and their requirements as follows. The unit falls under category II.
Measurement category IV is for measurement performed at the source of a
low-voltage installation.
Measurement category III is for measurement performed in a building
installation.
Measurement category II is for measurement performed on circuits directly
connected to a low voltage installation.
Measurement category I is for measurements performed on circuits not directly
connected to Mains.