User manual

91
7. Press Slope repeatedly to select the trigger slope,
which also appears at the bottom of the display.
Range Rising edge, falling edge
Slope
8. Press Coupling repeatedly to select the trigger
coupling.
Range DC, AC
Coupling
AC
9. Press Rejection to select the frequency rejection
mode.
Range LF, HF, Off
Rejection
Off
10. Press Noise Rej to turn the noise rejection on or off.
Range On, Off
Noise Rej
Off
11. Press Previous menu to go back to the previous
menu.
Previous
Menu
Manually triggering the signal
Note This section describes how to manually trigger the input signals when the oscil-
loscope does not capture them. This section applies to the Normal and Single
trigger mode, since in the Auto trigger mode, the oscilloscope keeps updating the
input signal regardless of the trigger conditions.