Data Sheet
TECHNICAL DATA
Display Max. 100,000 counts
Measuring rate 3 measuring operations/second
Measuring line length approx. 90 cm each
Measuring impedance 10MΩ (V range), <100pF
Operating voltage 4 mignon batteries (type AA)
Working conditions 11 to 30°C (<80%rF), >30 to 40°C (<75%rF),
>40 to 50°C (<45%rF)
Operating altitude max. 2,000 m
Storage temperature -20°C to +60°C
Weight approx. 620 g
Dimensions (LxWxH) 212 x 98 x 52 mm
Over-voltage category CAT III 1,000 V, CAT IV 600 V, contamination degree 2
Crest factor max 3 (CF 1.4 to 2.0 +1%; CF 2.0 to 2.5 +2.5%; CF 2.5 to 3.0
+4.0%)
Direct Voltage
Range Accuracy
100.000 mV ± (0.03 % + 4)
1000.000 mV ± (0.026 % + 20)
10.000 mV
± (0.02 % + 20)
100.000 mV
1000.000 mV
Alternating Voltage
Range
Accuracy
100.000 mV ± (1 % + 50) at 40 Hz
~
65 Hz
1000.000 mV
± (1.95 % + 50) at 66 Hz
~
1 kHz
± (3.9 % + 50) at 1.01 kHz
~
3 kHz
10.000 mV ± (1.3 % + 50) at 40 Hz
~
45 Hz
100.000 mV ± (0.52 % + 50) at 46 Hz
~
65 Hz
1000.000 mV ± (1.3 % + 50) at 66 Hz
~
1 kHz
10.000 mV ± (2.6 % + 50) at 1.01 kHz
~
10 kHz
100.000 mV
± (3.9 % + 50) at 10.01kHz
~
20kHz
± (6.5 % + 50) at 20.01kHz
~
50kHz
± (13 % + 50) at 50.01kHz
~
100kHz
Direct current
Range Accuracy
10.000 mV
± (0.13 % + 40)
100.000 mV
10.000 mV ± (0.13 % + 80)