Information
at different measurement mode.
Msg[5]:
Value DCV ACV DCmV OHM CAP DCuA DCmA DCA ACuA ACmA ACA FREQ
0x30 6V 6V 600mV 600Ω 60nF 600uA 60mA 10A 600uA 60mA 10A 60Hz
0x31 60V 60V 6KΩ 600 nF 6000uA 600mA 6000uA 600mA 600Hz
0x32 600V 600V 60KΩ 6000 nF 6kHz
0x33 1000V 1000V 600KΩ 60uF 60kHz
0x34 6MΩ 600μF 600kHz
0x35 60MΩ 6000μF 6MHz
0x36 60mF 60MHz
0x37
Display 1: Msg[6] – Msg[12]
7 bytes are used to represent the value shown on the Display 1 of the VC890 device. Please decode this value at
ASIC-II string on the PC application.
Display 2: Msg[13] – Msg[20]
8 bytes are used to represent the time.Please decode this value at ASIC-II string on the PC application.
Display 3: Msg[21] – Msg[30]
10 bytes are used to represent the date.Please decode this value at ASIC-II string on the PC application.
Display 4: Msg[31] – Msg[38]
8 bytes are used to represent the value shown on the Display 4. Please decode this value at ASIC-II string on the PC
application.
Display 5: Msg[39] – Msg[46]
8 bytes are used to represent the value shown on the Display 5. Please decode this value at ASIC-II string on the PC
application.
Second frequency Unit: Msg[47] – Msg[49]
3 bytes are used to represent the value shown on frequency Unit. Please decode this value at ASIC-II string on the PC
application.
Display 6: Msg[50] – Msg[53] :
4 bytes are used to represent the value shown on the Display 6. Please decode this value at ASIC-II string on the PC
application.
Display Bar: Msg[54] – Msg[55] :
2 bytes are used to represent the value shown on bar.
Msg[54]:
tens digit
Msg[55] :a digit