Datasheet
Impedance
Range Accuracy Resolution
400 Ω ±(1,0% + 10) with REL function 0,01 Ω
4 kΩ
±(0,6% + 10)
0,1 Ω
40 kΩ 10 Ω
400 kΩ 100 Ω
4 MΩ ±(1,2% + 10) 1 kΩ
40 MΩ ±(2% + 5) 10 kΩ
Overload protection 1000 V
Capacity
Range Accuracy Resolution
40 nF ±(2,5% + 20) 1 pF
400 nF
±(2% + 20)
10 pF
4 μF 100 pF
40 μF 1 nF
400 μF 10 nF
4000 μF ±(5% + 20) 100 nF
40 mF Not specified 1 μF
Overload protection 1000 V
Frequency
Range Accuracy Resolution
10 Hz - 40 MHz ±(0,02% + 8) 0,001 Hz - 0,001 MHz
400 MHz Not specified 0,01 MHz
Overload protection 1000 V
Sensitivity (10 Hz – 10 MHz): 200 mV; Amplitude max. 30 Veff (ms)
Sensitivity (10 MHz – 40 MHz): 400 mV; Amplitude max. 30 Veff (ms)
Duty-Cycle (pulse-break ratio)
Range Accuracy Resolution
5Hz ~ 2kHz
(10%~90%)
±(1,2% + 30) 0,01%
Overload protection 1000V
Temperature
Range Accuracy Resolution
-40 to +40 °C ±(3% + 20)
0,1 °C+40 to +400 °C ±(2% + 20)
+400 to +1000 °C ±2,5%
-40 to +32 °F ±(2,5 % + 40)
0,1 °F+32 to +752 °F ±(1,5 % + 40)
+752 to +1832 °F ±2,5 %
Diode test
Test voltage Resolution
2,73 V 0,0001 V
Overload protection: 1000 V