Data Sheet
GENERAL SPECIFICATIONS
SYMBOL: 6000 counts MEASURING RATE: Approx. 3 measuring operations/second MEASURING LINE LENGTH:
Approx. 90 cm each WORKING CONDITIONS: 0 to 30°C (<75 %rF), 30 to 40°C (<50 %rF) OPERATING ALTITUDE:
max. 2,000 m STORAGE TEMPERATURE: -10°C to +50°C BASIC ACCURACY: ± 5 % VOLTAGE SUPPLY: 9 V block
battery OVER-VOLTAGE CATEGORY: CAT III 1000 V, CAT IV 600 V, contamination degree 2 WEIGHT: Approx. 380 g
DIMENSIONS (W X H X D): 185 x 91 x 43 (mm)
TECHNICAL DATA
Functions Range Resolution
Accuracy
±(a% reading + b digits)
Overload
protection
DCV
600 mV 0.1 mV ±(0.5 % + 3)
1000 V
Impedance: 10 MΩ
6 V / 60 V / 600 V 1 mV / 10 mV / 100 mV ±(0.5 % + 2)
1000 V 1 V ±(0.8 % + 3)
ACV
1
6 V / 60 V / 600 V 1 mV / 10 mV / 100 mV ±(0.8 % + 3)
Frequency range 45 – 400 Hz;
Overload protection 750 V.
750 V 1 V ±(1.0 % + 5)
DCA
600 μA / 6 mA / 60 mA / 600 mA 0.1
μA / 1 μA / 10 μA/ 100 μA ±(0.8 % + 3)
Fuse, measuring time limit > 5 A:
max. 10 s with 10 min break.
6 A / 10 A 1 mA / 10 mA ±(1.2 % + 5)
ACA
2
600 μA / 6 mA 0.1 μA / 1 μA ±(1.0 % + 2)
Fuse, measuring time limit > 5 A:
max. 10 s with 10 min break.
Frequency range 45 – 400 Hz;
Overload protection 750 V.
60 mA / 600 mA 10
μA / 100 μA ±(1.2 % + 3)
6 A / 10 A 1 mA / 10 mA ±(1.5 % + 5)
Ω
600 Ω 0.1 Ω
±(0.8 % + 3)
w/ REL function
Overload protection 1000 V;
Measuring voltage: Approx.
0.4 V.
6 k
Ω / 60 kΩ / 600 kΩ 1 Ω / 10 Ω / 100 Ω ±(0.8 % + 2)
6 M
Ω 1 kΩ
±(1.2 % + 2)
≥ 20 MΩ:
±(1.5 % + 5)
60 M
Ω 10 kΩ
Capacity 40 nF 0.01 nF
±(3.0% + 5)
with RELfunction
Overload protection 1000 V
400 nF 0.1 nF
4
μF 1 nF
40
μF 10 nF ±(3.0 % + 5)
400
μF 100 nF ±(4.0 % + 5)
4000
μF 1 μF Not Specified
Frequency /
Duty Cycle
3
10 Hz - 10 MHz 0.001 Hz – 0.01 MHz ±(0.1 % + 3)
Overload protection 750 V
0.1 – 99.9 % 0.1 %
Not specified
Diode test Approx. 3.7 V 0.001 V
NA
Overload protection 1000 V
Acoustic
continuity
tester
<30 Ω continuous sound, test voltage: approx. 0.65 V/DC;.
Overload protection 1000 V
1. Effective average (RMS) at sinus voltage. Permissible display error with an open measuring input: 2 counts. Perissible display error with a short-circuited measuring input: 20 counts.
2. Effective average (RMS) at sinus voltage
3. Sensitivity: 200 mV; Amplitude max. 5 Veff (>5 V with adjustable impedance)
PACKAGE CONTENT
Unit // Batteries // Operating instructions