Datasheet

Measuring function AC + DC voltage
Range Accuracy Resolution Frequency range
4 V
±(1 % + 80 digits)
0.0001 V
45 – 1,000 Hz
±(3 % + 40 digits) 1,000 – 10,000 Hz
±(6 % + 40 digits) 10,000 – 35,000 Hz
40 V
±(1 % + 80 digits)
0.001 V
45 – 1,000 Hz
±(3 % + 40 digits) 1,000 – 10,000 Hz
±(6 % + 40 digits) 10,000 – 35,000 Hz
400 V
±(1 % + 80 digits)
0.01 V
45 – 1,000 Hz
N/A 1,000 – 10,000 Hz
N/A 10,000 – 35,000 Hz
600 V
±(1 % + 80 digits)
0.1 V
45 – 1,000 Hz
N/A 1,000 – 10,000 Hz
N/A 10,000 – 35,000 Hz
Overload protection: 1000 V; Impedance: about 10 MΩ
Direct current
Range Accuracy Resolution
400 µA ±(0.3 % + 10 digits) 0.01 µ A
4000 µA ±(0.5 % + 10 digits) 0.1 µA
40 mA 0.001 mA
400 mA 0.01 mA
10 A ±(1.5 % + 20 digits) 0.001 A
Alternating current
Range Accuracy Resolution Frequeny range
400 µA
±(0.6 % + 40 digits)
0.01 µ A
45 – 1,000 Hz
±(1.2 % + 40 digits) 1,000 – 10,000 Hz
400 µA
±(0.6 % + 40 digits)
0.1 µA
45 – 1,000 Hz
±(1.2 % + 40 digits) 1,000 – 10,000 Hz
400 mA
±(0.6 % + 40 digits)
0.001 mA
45 – 1,000 Hz
±(1.2 % + 40 digits) 1,000 – 10,000 Hz
400 mA
±(0.6 % + 40 digits)
0.01 mA
45 – 1,000 Hz
±(1.2 % + 40 digits) 1,000 – 10,000 Hz
10 A
±(2 % + 40 digits)
0.001 A
45 – 1,000 Hz
±(4 % + 40 digits) 1,000 – 10,000 Hz
Overload protection: Fuses, 500 mA (µ AmA) / 10 A(10 A); measuring time limit max. 10 s with
10 min break; Overload protection: 1000 V; TrueRMS in the measuring range of 10 – 100 %
Impedance
Range Accuracy Resolution
400 Ω ±(1.0 % + 10 digits) with REL
function
0.01 Ω
4 kΩ ±(0.6 % + 10 digits) 0.0001 kΩ
40 kΩ 0.01 kΩ
400 kΩ 0.1 kΩ
4 MΩ ±(1.2 % + 10 digits) 0.001 MΩ
40 MΩ ±(2 % + 5 digits) 0.01 MΩ
Overload protection: 1000 V