Alternating current
Range Accuracy Resolution Crest factor max.
600.0 µA
±(1.3% + 6)
0.1 µA 3.0
6000 µA 0.001 mA 3.0
60.00 mA 0.01 mA 3.0
600.0 mA 0.1 mA 3.0
6.000 A
±(1.5% + 8)
0.001 A 3.0
10.00 A 0.01 A 3.0
Specied measured range: 5 - 100 % of the measured range
Permissible display with a short-circuited measuring input: <3 counts
Frequency range 45 - 400 Hz
Measuring converter AC-A: AC-coupled, real effective value (True RMS)
Additional tolerance at the following crest factor of a non-sine-shaped measuring signal:
CF 1.0 – 2.0: +3% for accuracy
CF 2.0 – 2.5 +5% for accuracy
CF 2.5 – 3.0 +7% for accuracy
Overload protection 10 A: Fuse; F10 A H 600 V; measuring time limit >5 A: max. 10 s with
15 min break
Overload protection µA/mA: PTC fuse: 600 mA F 600 V
Resistance
Range Accuracy Resolution
600.0 Ω ±(1.2% + 6) with REL function 0.1 Ω
6.000 kΩ
±(1.2% + 6)
0.001 kΩ
60.00 kΩ 0.01 kΩ
600.0 kΩ 0.1 kΩ
6.000 MΩ ±(1.5% + 6) 0.001 MΩ
60.00 MΩ ±(2.0% + 6) 0.01 MΩ
Overload protection 600 V PTC; measuring voltage: approx. 1 V