User manual

53
Alternating voltage
Range Accuracy Resolution Crest factor max.
60.00 mV ±(1.5% + 4) 0.01 mV 3.0
600.0 mV 0.1 mV 3.0
6.000 V
±(1,3% + 4)
0.001 V 3.0
60.00 V 0.01 V 3.0
600.0 V ± (1,5% + 4) 0.1 V
3.0
1.5 (at measuring range end
value)
6 - 600 V LoZ ±(2.6% + 6)
Specied measured range: 5 - 100 % of the measuring range; LoZ = low impedance 400 kΩ
Permissible display with a short-circuited measuring input: <3 counts
Frequency range 45 – 400 Hz; Overload protection 600 V
Measuring converter AC-V: AC-coupled, real effective value (True RMS), impedance: >10 MΩ, 100 pF
Additional tolerance at the following crest factor of a non-sine-shaped measuring signal:
CF 1.0 – 2.0 +3% for accuracy
CF 2.0 – 2.5 +5% for accuracy
CF 2.5 – 3.0 +7% for accuracy
Direct current
Range Accuracy Resolution
600.0 µA
±(1.0% + 7)
0.1 µA
6000 µA 1 µA
60.00 mA 0.01 mA
600.0 mA 0.1 mA
6.000 A
±(1.2% + 6)
0.001 A
10.00 A 0.01 A
Overload protection 10 A: Fuse; F10 A H 600 V; measuring time limit >5 A: max. 10 s with 15 min
break
Overload protection µA/mA: PTC fuse: 600 mA F 600 V