5.15.7 [MkrΔ->CF]
Changes the center frequency of the analyzer to the frequency
difference between the two markers.
5.16
【
Marker Fctn
】
Marker Function
Executes specific marker soft menu.
5.16.1 [Function Off]
Turn off marker measurement function.
5.16.2 [NdB On Off]
Enables the N dB BW measurement or sets the value of N. The N
dB BW denotes the frequency difference between points that are
located on both sides of the current marker while the amplitude
falls off (N<0) or rises (N>0) N dB separately,
Key points:
When the measurement starts, the analyzer will search the
two points which are located at both sides of the current point
and are N dB amplitudes smaller or greater than the current
point, and display the frequency difference between the two
points.
You can use the numeric keys, knob or direction keys to
modify the value of N, 3 at default.
5.16.3 [Marker Noise On Off]
Turn on or off the frequency noise function. The function of marking
noise is applied to the selected cursor, and then the noise Power
Spectral Density at the cursor is read. When turned on, the