will be.
Changing the points would also influence other system
parameters, thus the instrument restarts the sweep and
measurement cycle.
You can modify this parameter using the numeric keys, knob or
direction keys.
5.10
【
Trig
】
Sets the trigger type and other associated parameters, menu
includes [Auto Run] and [Video].
5.10.1 [Auto Run]
Set the trigger mode to the free trigger mode so that the scan
trigger is as fast as possible with the spectrum analyzer. It meets
the trigger conditions at any time, that is, continue to generate a
trigger signal.
5.10.2 [Video]
This indicates a trigger signal will be generated when the system
detects a video signal in which the voltage exceeds the specified
video trigger level.
5.11
【
Source
】
When the Source is On, a signal with the same frequency of the
current sweep signal will be output from the GEN OUTPUT 50Ω
terminal on the front panel. Press the key will pop out related menu
includes [Tracking Source►], [Tracking Source On Off], [Output
Power], [Network Measure►].The tracking source is turned off in
the power-on and reset states.
5.11.1 [Track Gen►]
To configure tracking source.