5.2.3 [Zero Span]
Sets the span of the analyzer to 0 Hz. Both the start and stop
frequencies will equal the center frequency and the horizontal axis
will denote time. The analyzer here is measuring the time domain
characteristics of amplitude, located at the corresponding
frequency point. This will help to observe the signal (especially for
modulated signal) at time domain.
5.2.4 [Last Span]
Changes the span to the previous span setting.
5.3
【
AMPTD
】
Amplitude
Sets the amplitude parameters of the analyzer. Through these
parameters, signals under measurement can be displayed at an
optimal view with minimum error. The pop out amplitude menu
includes [Ref Level], [Attenuation Auto Manual], [Scale/Div],
[Scale Type Lin Log], [Ref Offset], [Ref Unit], and [Preamplifier
On Off].
5.3.1 [Ref Level]
Activate reference level function and sets the maximum power or
voltage for display window.
Key points:
This value is affected by a combination of maximum mixing level,
input attenuation, and preamplifier. When you adjust it, the input
attenuation is adjusted under a constant max mixing level,
meeting:
L
Ref
− a
RF
+ a
PA
≤
L
mix
(5-3)
L
Ref
,
a
RF
,
a
PA
and
L
mi
x
denotes the reference level, the input
attenuation, the preamplifier, and the max mixing level,
respectively.
28