Instructions
SCPI Command Set
45
1
Window sweep is on.
Example
:PASSFAIL:WIN:SWEEP:STAT 1
[:SENSe]:POWer[:RF]:ATTenuation
Description
Sets or queries the input attenuation.
Syntax
Query Syntax
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>
[:SENSe]:POWer[:RF]:ATTenuation?
Parameter/
Return parameter
<rel_ampl>
<NR1> 0 dB to 40 dB
Example
:POW:ATT 10 dB
[:SENSe]:POWer[:RF]:ATTenuation:AUTO
Description
Sets or queries whether the automatic input attenuation is on/off.
Syntax
Query Syntax
[:SENSe]:POWer[:RF]:ATTenuation:AUTO {OFF|ON|0|1}
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?
Parameter
0
1
OFF
ON
Turns automatic input attenuation off.
Turns automatic input attenuation on.
Turns automatic input attenuation off.
Turns automatic input attenuation on.
Return parameter
0
1
Automatic input attenuation is off.
Automatic input attenuation is on.
Example
:POW:ATT:AUTO ON
[:SENSe]:POWer[:RF]:GAIN[:STATe]:AUTO
Description
Turns the preamplifier on/off or queries its state.
Syntax
Query Syntax
[:SENSe]:POWer[:RF]:GAIN[:STATe]:AUTO {OFF|ON|0|1}
[:SENSe]:POWer[:RF]:GAIN[:STATe]:AUTO?
Parameter
0
1
OFF
ON
Turns the preamplifier off.
Turns the preamplifier on.
Turns the preamplifier off.
Turns the preamplifier on.
Return parameter
0
1
The preamplifier is off.
The preamplifier is on.
Example
:POW:GAIN:AUTO ON