Instructions

SCPI Command Set
42
Syntax
Query Syntax
[:SENSe]:PASSFAIL:LINELimit:X:OFFSET <freq>
[:SENSe]:PASSFAIL:LINELimit:X:OFFSET?
Parameter
<freq>
<NRf> Frequency in Hz
Return parameter
<NR3>
Hz
Example
:PASSFAIL:LINEL:X:OFFSET 1.0e+6
[:SENSe]:PASSFAIL:LINELimit:Y:OFFSET
Description
Sets or queries the amplitude offset (Shift Y) of the upper and lower limit line.
Syntax
Query Syntax
[:SENSe]:PASSFAIL:LINELimit:Y:OFFSET <ampl>
[:SENSe]:PASSFAIL:LINELimit:Y:OFFSET?
Parameter
<ampl>
<NRf> Amplitude in dBm
Return parameter
<NR3>
dBm
Example
:PASSFAIL:LINEL:Y:OFFSET -15
[:SENSe]:PASSFAIL:WINdow:AMPl:LOWer
Description
Sets or queries the lower amplitude of amplitude line in window testing.
Syntax
Query Syntax
[:SENSe]:PASSFAIL:WINdow:AMPl:LOWer <ampl>
[:SENSe]:PASSFAIL:WINdow:AMPl:LOWer?
Parameter
<ampl>
Amplitude in dBm
Return parameter
<NR3>
dBm
Example
:PASSFAIL:WIN:AMPl:LOW -20
[:SENSe]:PASSFAIL:WINdow:AMPl:UPper
Description
Sets or queries the upper amplitude of amplitude line in window testing.
Syntax
Query Syntax
[:SENSe]:PASSFAIL:WINdow:AMPl:UPper <ampl>
[:SENSe]:PASSFAIL:WINdow:AMPl:UPper?
Parameter
<ampl>
Amplitude in dBm
Return parameter
<NR3>
dBm
Example
:PASSFAIL:WIN:AMPl:UP -10
[:SENSe]:PASSFAIL:WINdow:AMPt:STATe
Description
Turns the amplitude line of window testing on/off or queries its state.
Syntax
[:SENSe]:PASSFAIL:WINdow:AMPt:STATe {OFF|ON|0|1}