Instructions

SCPI Command Set
20
OFF
ON
Return parameter
0
1
Example
:DISP:WIN:TRAC:Y:GAU ON
:DISPlay:WINdow:TRACe:Y[:SCALe]:PDIVision
Description
Sets or queries the Y-axis scale/div when the amplitude scale is logarithmic.
Syntax
Query Syntax
:DISPlay:WINdow:TRACe:Y[:SCALe]:PDIVision <rel_ampl>
:DISPlay:WINdow:TRACe:Y[:SCALe]:PDIVision?
Parameter/
Return parameter
<rel_ampl>
<NR1> 1 dB to 255 dB
Example
:DISP:WIN:TRAC:Y:PDIV 10
:DISPlay:WINdow:TRACe:Y[:SCALe]:RLEVel
Description
Sets or queries the Y-axis reference level. The units depend on the scale type
(logarithmic/linear).
Syntax
Query Syntax
:DISPlay:WINdow:TRACe:Y[:SCALe]:RLEVel <ampl>
:DISPlay:WINdow:TRACe:Y[:SCALe]:RLEVel?
Parameter/
Return parameter
<ampl>
<NRf> in current active unit
<NR3>
Current active unit
Example
:DISP:WIN:TRAC:Y:RLEV -10
:DISPlay:WINdow:TRACe:Y[:SCALe]:RLEVel:OFFSet
Description
Sets or queries the Y-axis reference level offset.
Syntax
Query Syntax
:DISPlay:WINdow:TRACe:Y[:SCALe]:RLEVel:OFFSet <rel_ampl>
:DISPlay:WINdow:TRACe:Y[:SCALe]:RLEVel:OFFSet?
Parameter
<rel_ampl>
<NRf> dB
Return parameter
<NR3>
Example
:DISP:WIN:TRAC:Y:RLEV:OFFS -5.0e+1 dB
:DISPlay:WINdow:TRACe:Y[:SCALe]:SPACing
Description
Sets or queries the type of scale: logarithmic or linear.
Syntax
:DISPlay:WINdow:TRACe:Y[:SCALe]:SPACing {LINear|LOGarithmic}