Instructions
SCPI Command Set
13
Parameter
<n>
<freq>
<NR1> Marker number 1 to 5
Frequency in GHz, MHz, kHz, Hz. The default unit is Hz.
Return parameter
<freq>
Frequency in Hz.
If the specified marker is not active, returns ERR.
Example
:CALC:MARK2:X 300MHz
query
:CALC:MARK2:X?
returns
300000000Hz
:CALCulate:MARKer<n>:Y?
Description
Returns the marker’s vertical position. In delta marker mode, this command
returns the amplitude difference between the markers.
Query Syntax
:CALCulate:MARKer<n>:Y?
Parameter
<n>
<NR1> Marker number 1 to 5
Return parameter
<ampl>
Power or voltage.
If the specified marker is not active, returns ERR.
Example
:CALC:MARK3:Y?
returns
9.8dBm
:CALCulate:NETMeasure:POWer
Description
Sets or queries the network measurement output power level.
Syntax
Query Syntax
:CALCulate:NETMeasure:POWer <ampl>
:CALCulate:NETMeasure:POWer?
Parameter
<ampl>
<NRf> Power or voltage, -30dBm to 0dbm
Return parameter
<NR2>
Example
:CALC:NETM:POW -10
:CALCulate:NTDate:NORMalize
Description
Turns the tracking generator normalization on/off or queries its state.
Syntax
Query Syntax
:CALCulate:NTDate:NORMalize {OFF|ON|0|1}
:CALCulate:NTDate:NORMalize?