Instructions
4.Advanced User Guidebook 
87 
2.  In the bottom menu, select Source as CH1. 
3.  Select the first menu item in the bottom menu, the Type menu will display at 
the right of the screen. In the right menu, select Time&Voltage for Type, 
two blue dotted lines displayed along the horizontal direction of the screen, 
two  blue  dotted lines displayed along the vertical  direction of the screen. 
Cursor measure window  at the left bottom of the screen  shows  the cursor 
readout. 
4.  In the bottom menu, select Line Type as Time to make the vertical cursors 
active. If the Line in the bottom menu is select as a, turn the M knob to move 
line a to the right or left. If b is selected, turn the M knob to move line b. 
5.  In  the bottom menu, select  Line Type  as  Voltage  to make the horizontal 
cursors active. Select Line in the bottom menu as a or b, turn the M knob to 
move it. 
6.  Push the Horizontal HOR button to enter wave zoom mode. In the bottom 
cursor menu, select Window  as  Main  or  Extension  to  make the cursors 
shown in the main window or zoom window. 
Figure 4-17 Time&Voltage Cursor Measurement   
Auto Cursor 
For the AutoCursr type, the horizontal cursors are set as the intersections of the 
vertical cursors and the waveform. 