Instructions
4.Advanced User Guidebook 
80 
4.  Under the status of stop, data comparing will stop, and when it goes on running, 
the number of Pass/Fail will increase from the former number, not from zero. 
5.  When  the waveform playback mode is on,  Pass/Fail  is used to test the 
played-back waveform specially. 
●  Output 
Push the Utility button, select Function in the bottom menu, select Output in the left 
menu. 
Output menu item in the bottom menu sets the output type of Trig Out(P/F) port in 
Figure 3-3 Rear Panel on P6. In the bottom menu, select Output. The description of 
Output menu is shown as the follows: 
Device and Print Setup menu items set the print output, refer to "How to Print the 
Screen Image" on page 93. 
●  LAN Set 
Using the LAN port, the oscilloscope can be connected with a computer.   
●  Update 
Use the front-panel USB port to update your instrument firmware using a USB 
memory device. Refer to "How to Update your Instrument Firmware" on page 80. 
●  DAQ 
You can use the multimeter data recorder to record the measurements when 
measuring current/voltage by multimeter (optional). Refer to "Multimeter Recorder" 
on page 107. 
●  FRA (Frequency Response Analysis) 
If there is a built-in arbitrary function generator (optional), you can use the frequency 
response analysis. Refer to "Frequency Response Analysis" on page 101. 
How to Update your Instrument Firmware 
Use the front-panel USB port to update your instrument firmware using a USB 
Function 
Menu 
Setting  Description 
Type 
Trig level 
Output trig signal synchronously 
Pass/fail 
Output High Level when Pass , and Low Level when Fail 
AG 
Output 
CH2  Output of  dual-channel  waveform generator 
(optional) 